Summary0006322: Aurora code fails for 5D sensors
DescriptionIn the new tracker architecture, the aurora code fails to recognize two 5D sensors attached to a splitter
Steps To ReproducePlug two 5D sensors into the same port through spliter
Additional InformationSpecial Instructions on split port from Aurora API Guide:
For a split port on a dual 5DOF tool, the first PHSR sent will report only one port handle. After the port handle is initialized, it is assigned to channel 0. You must then use PHSR again to assign a port handle to channel 1. The port handle for channel 1 is initialized automatically.
TagsNo tags attached.
Attached Files

 Relationships

  Notes
(0014034)
Patrick Cheng (developer)
2008-11-05 14:36

Add additional split port detecting command in AuroraTracker::ValidateSpecifiedFrequency

 Issue History
Date Modified Username Field Change
2008-02-12 08:56 Andinet New Issue
2008-02-12 08:56 Andinet Status new => assigned
2008-02-12 08:56 Andinet Assigned To => Patrick Cheng
2008-02-12 08:56 Andinet Assigned To Patrick Cheng => Andinet
2008-11-05 14:33 Patrick Cheng Steps to Reproduce Updated
2008-11-05 14:33 Patrick Cheng Additional Information Updated
2008-11-05 14:34 Patrick Cheng Status assigned => confirmed
2008-11-05 14:34 Patrick Cheng Status confirmed => assigned
2008-11-05 14:34 Patrick Cheng Assigned To Andinet => Patrick Cheng
2008-11-05 14:36 Patrick Cheng Note Added: 0014034
2008-11-05 14:36 Patrick Cheng Status assigned => resolved
2008-11-05 14:36 Patrick Cheng Resolution open => fixed
2008-11-05 14:38 Patrick Cheng Status resolved => closed