MantisBT - IGSTK
View Issue Details
0006322IGSTKIGSTKpublic2008-02-12 08:562008-11-05 14:38
Andinet 
Patrick Cheng 
normalminoralways
closedfixed 
0006322: Aurora code fails for 5D sensors
In the new tracker architecture, the aurora code fails to recognize two 5D sensors attached to a splitter
Plug two 5D sensors into the same port through spliter
Special Instructions on split port from Aurora API Guide:
For a split port on a dual 5DOF tool, the first PHSR sent will report only one port handle. After the port handle is initialized, it is assigned to channel 0. You must then use PHSR again to assign a port handle to channel 1. The port handle for channel 1 is initialized automatically.
No tags attached.
Issue History
2008-02-12 08:56AndinetNew Issue
2008-02-12 08:56AndinetStatusnew => assigned
2008-02-12 08:56AndinetAssigned To => Patrick Cheng
2008-02-12 08:56AndinetAssigned ToPatrick Cheng => Andinet
2008-11-05 14:33Patrick ChengSteps to Reproduce Updated
2008-11-05 14:33Patrick ChengAdditional Information Updated
2008-11-05 14:34Patrick ChengStatusassigned => confirmed
2008-11-05 14:34Patrick ChengStatusconfirmed => assigned
2008-11-05 14:34Patrick ChengAssigned ToAndinet => Patrick Cheng
2008-11-05 14:36Patrick ChengNote Added: 0014034
2008-11-05 14:36Patrick ChengStatusassigned => resolved
2008-11-05 14:36Patrick ChengResolutionopen => fixed
2008-11-05 14:38Patrick ChengStatusresolved => closed

Notes
(0014034)
Patrick Cheng   
2008-11-05 14:36   
Add additional split port detecting command in AuroraTracker::ValidateSpecifiedFrequency